Atomic
Absorption Spectrophotometers
Infra-Red Spectrophotometers
UV/Visible Spectrophotometers
Gas Chromatographs
Dedicated Analyzers
High Performance Liquid Chromatographs
Accessories & Supplies

|
| 
| Part Number: |
Description: |
| |
Atomic Absorption Spectrophotometer |
Detailed
Product Information
The
205 Advantage
The Buck Scientific 205 AAS uses a unique burner
system to get more sample to the flame, which
results in more signal and better sensitivity.
We use an impact bead nebulizer to get an efficient
generation of a fine aerosol of sample solution.
The short path spray chamber permits difficult
solutions containing organics or high dissolved
solids to be aspirated with superb precision
and minimal instabilities. To take advantage
of this enhanced signal, we designed an optical
layout that provides the highest throughput
of any commercially available spectrophotometer.
Our very simple straight line optical path
uses a monochromator with a single reflector
mirror.
This cuts down on the amount of light loss
often
associated with models having 10 or more mirrors.
Because of this the Buck 205 AAS can detect
far-UV elements, such as arsenic and selenium,
with ten times greater sensitivity.
| • |
Internal microprocessor and software can
identify samples, perform statistical analysis
and graphically display calibration curves. |
| • |
Dual lamp power, one for warm up
|
| • |
High throughput "Stable Beam"
optical system |
| • |
Modulated HCL power supply to cancel DC
emission from flame |
| • |
Digital
readout for absorbance, emission, direct
concentration, lamp currents and PMT voltage |
| • |
Autozero,
curve correction, integration and high gain
damping features |
| • |
Bandpass
resolution from 0.2 - 0.7 - 2.0nm |
| • |
Burner
safety interlock system for nitrous oxide/acetylene
flame operation |
| • |
High
sensitivity detection using quartz PMT
detector |
| • |
Analog
and digital output for chart recorders/PC
data systems |
| • |
Compact
design saves bench space |
Wavelength
Range: |
190
to 900nm |
Burner/
Nebulizer: |
Polypropylene
spray chamber with premix burner and
high
efficiency adjustable nebulizers (SS or
Pt/Ru), Titanium burner heads for Air/Acetylene,
Argon/Hydrogen and nitrous Oxide/Acetylene
operation |
| Accuracy: |
±0.2nm |
Integration: |
User
selectable times from 0.5 to 10 seconds
for Stabilized readings. |
| Precision: |
±0.1nm |
Calibration: |
Automatic,
weighted least squares fit to 1st, 2nd,
3rd, or 4th order functions, up to 8
points. |
| Monochromator
/ Optics: |
- 250mm
Ebert Mount
- 600
lines/mm high efficiency grating
- 0.2,
0.7, 2.0nm bandpass to cover analytical
wavelengths for use on atomic absorption
and emission measurements
|
Display: |
16
line backlit liquid crystal display for
all text & graphics |
| Power
Supply: |
100-240V.A.C.
operating range, 50/60HzAC,<100 watts
|
Output
Modes: |
LCD
Display, IEEE Parallel port for dot matrix
printer, RS-232 Serial Port |
| Hollow
Cathode |
|
Dimensions: |
39"
x 11" x 12" |
| Lamp: |
Dual
HCL modulated power supply, 0 to 25mA peak |
Weight: |
50
lbs. (80 lbs. shipping) |
| |
|
|
|
|
| |
|