Thermal Conductivity Detector (TCD)
View Full-Size Image


Thermal Conductivity Detector (TCD)

Price per Unit (piece): $2,059.00

Request Information / Request for Quote

  • Universal detector
  • Detects from 100% down to 200-500 ppm
  • Easily replaceable filaments
  • Thermostatted up to 275° C
Our TCD is equipped with user-replaceable filaments, so it can be quickly returned to service in the event of a burn-out.

Because it detects all molecules, the Thermal Conductivity Detector is commonly used for fixed gas analysis (O2, N2, CO, CO2, H2S, NO, etc.) where the target analytes do not respond well on other, more sensitive detectors. The TCD can detect concentrations from 100% down to around 100 ppm on a flat baseline with sharp peaks. Where the peak is broad or the baseline is not flat, detection limits of 300 ppm are more realistic. For lower detection limits, the HID detector may be more suitable for inorganics, and the FID detector provides 1ppm detection for hydrocarbon species.

The TCD consists of four tungsten-rhenium filaments in a Wheatstone bridge configuration. Electric current flows through the filaments, causing them to heat up. Carrier gas (usually helium, which has very high thermal conductivity) flows across the filaments, removing heat at a constant rate. Two of the filaments are exposed only to carrier gas (reference), and two are exposed to the carrier/sample flow. When a sample molecule with lower thermal conductivity than the carrier gas exits the column and flows across the two sample filaments, the temperature of the filaments increases. This temperature increase unbalances the Wheatstone bridge and generates a peak as sample molecules transit through the detector.

A filament protection circuit prevents filament damage by disabling the current if carrier pressure is not detected by the GC, but cannot prevent filament damage in all circumstances. The TCD is equipped with user-replaceable filaments in the event of a burn-out.

This chromatogram shows 0.5mL 10,000 ppm (1%) CO2 separated from air using a 1-meter Silica Gel packed column at 80°C.








Customer Reviews:

There are yet no reviews for this product.

Write a review for this product!


First: Rate the product. Please select a rating between 0 (poorest) and 5 stars (best).


Now please write a (short) review....(min. 1, max. 2000 characters)


Enter the code below:
Loading captcha image

Characters written:


View Cart

My Cart:
0 Items Order Total: $0.00
0 Product
$0.00



Free Shipping

for domestic orders over $500

 

Buck Scientific 2nd annual

Open House Training

September  4-6, 2012

Fill out the

Buck training application.pdf

and send it back to us

to reserve your spot!

Buck Scientific Privacy Policy

Upcoming Events

March 2012
Su Mo Tu We Th Fr Sa
     1 2 3
4 5 6 7 8 9 10
1112The Pittsburg Conference
13The Pittsburg Conference
14The Pittsburg Conference
15The Pittsburg Conference
16 17
18 19 20 21 22 23 24
25ACS Spring Meeting, San Diego CA
26ACS Spring Meeting, San Diego CA
27ACS Spring Meeting, San Diego CA
28ACS Spring Meeting, San Diego CA
29ACS Spring Meeting, San Diego CA
30 31
April 2012
Su Mo Tu We Th Fr Sa
1 2 3 4 5 6 7
8 9 10 11 12 13 14
15 1617Analitica Germany 2012
18Analitica Germany 2012
19Analitica Germany 2012
20Analitica Germany 2012
21
22 23 24 25 26 27 28
29 30     
May 2012
Su Mo Tu We Th Fr Sa
   1 2 3 4 5
6 7 8 9 10 11 12
13 14 15 1617 18 19
20 21 22 23 24 25 26
27 28 29 30 31  
June 2012
Su Mo Tu We Th Fr Sa
      1 2
3 4 5 6 7 8 9
10 11 12 13 14 15 16
17 18 19 20 21 22 23
24 25 26 27 28 29 30